二氧化氮氣體檢測管9DL 9DL
產品簡介
二氧化氮氣體檢測管9DL Nitrogen dioxide
產品詳細信息
二氧化氮氣體檢測管9DL Nitrogen dioxide
簡單介紹
二氧化氮氣體檢測管9DL Nitrogen dioxide,Measuring Range 0.01 to 3.0 ppm Sampling Hours 1 to 24 hours Detecting Limit 0.01 ppm Color Change White → Green Reaction Principle二氧化氮氣體檢測管價格
二氧化氮氣體檢測管9DL Nitrogen dioxide的詳細介紹
二氧化氮氣體檢測管9DL Nitrogen dioxide
Nitrogen dioxide No.9DL
NO2

| Performance |
|
Measuring Range |
0.01 to 3.0ppm |
|
Sampling Hours |
1 to 24 hours |
|
Detecting Limit |
0.01 ppm |
|
Color Change |
White→ Green |
|
Reaction Principle |
ABTS +NO2 → Green product |
| Coefficient ofVariation |
10%(for 0.01 to 3.0 ppm-hr) |
| ShelfLife | 1 Year |
| Correctionsfor temperature & humidity |
Unnecessary |
| Store the tubes inthe refrigerator to keep at 10ºC (50ºF) orbelow. | |
| Possible coexisting substances and theirinterferences |
|
Substance |
Concentration |
Interference |
Change color by itself |
|
Carbonmonoxiide |
10ppm |
No |
No discoloration (10 ppm) |
| Chlorine |
0.07 ppm |
No |
No discoloration(0.07 ppm) |
| Ozone |
0.045 ppm |
No |
No discoloration (0.045 ppm) |
| sulphur dioxide |
0.1 ppm |
|
No discoloration (0.1ppm) |
| Carbon dioxide |
600 ppm |
No |
No discoloration (600ppm) |
| Formaldehyde |
0.1ppm |
No discoloration (0.1ppm) |
| Calibration gasgeneration | Diffusion tubemethod | |
| TLV-TWA | TLV-STEL | Explosive range |
| 3 ppm | 5 ppm | - |